The development and validation of an UHPLC-MS/MS method for the rapid quantification of the antiretroviral agent dapivirine in human plasma


Background: Dapivirine is a non-nucleoside reverse transcriptase inhibitor designed to prevent HIV-1 viral replication and subsequent propagation. A sensitive method is required to quantify plasma concentrations to assess drug efficacy. Results: Dapivirine-spiked plasma was combined with acetonitrile containing deuterated IS and was processed for analysis. The method has an analytical measuring range from 20 to 10,000 pg/ml. For the LLOQ, low, mid and high QCs, intra- and inter-assay precision (%CV) ranged from 5.58 to 13.89% and 5.23 to 13.36%, respectively, and intra- and inter-day accuracy (% deviation) ranged from -5.61 to 0.75% and -4.30 to 6.24%, respectively. Conclusion: A robust...

To view this content, please register now for access

It's completely free